User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HEA 4934A/4935A DISTORTION CALIBRATION DATE: 20-Jul-93 AUTHOR: User Contributed REVISION: 1 F. Samonski 18-May-93 Added 4934A ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 71 CONFIGURATION: HP 8903B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J S M F W 1.002 ASK+ X 1.003 HEAD [27][91]1m {DISTORTION CALIBRATION} 1.004 DISP 1.004 DISP The HEA 8903A Distortion Analyzer must be connected 1.004 DISP to the Port 0 and be at address 28 for these tests. 1.005 DISP 1.005 DISP Remove all Connections from the UUT. 1.005 DISP 1.006 DISP Set UUT controls as follows: 1.006 DISP 1.006 DISP REV OFF 1.006 DISP DISPLAY TRMT 1.006 DISP SETUP TRMT 600 ohm ON 1.006 DISP MEASUREMENT MODE LEVEL FREQUENCY 1.006 DISP FREQUENCY READOUT 100 Hz 1.006 DISP OUTPUT LEVEL 0 dBm 1.007 DISP 1.007 DISP Connect the UUT TRMT/RCV phone jack to the 8903B 1.007 DISP HIGH INPUT, terminated with 600 ohms. Set the 8903B 1.007 DISP input switch to FLOAT. 1.007 DISP 1.008 IEEE [@8903][TERM LF] 1.009 IEEE [@8903]41.0SP 1.010 IEEE [@8903]M3LG 1.011 IEEE [@8903]T3T3RR[I] 1.012 MEME 1.013 MEME -55 1.014 MEME 1.015 MEMC -55 -55DB +100U -0U DIST 2.001 DISP 2.001 DISP Set the UUT FREQUENCY for A 100KHZ output. 2.002 IEEE [@8903]T3T3RR[I] 2.003 MEME 2.004 MEME -50 2.005 MEME 2.006 MEMC -50 -50DB +100U -0U DIST 3.001 DISP 3.001 DISP Set the UUT FREQUENCY for A 4KHZ output. 3.002 IEEE [@8903]T3T3RR[I] 3.003 MEME 3.004 MEME -55 3.005 MEME 3.006 MEMC -55 -55DB +100U -0U DIST 4.001 DISP 4.001 DISP Connect the UUT output through a 600 ohm load to the 4.001 DISP input of a 4KHz low pass filter. (KROHN-HITE 3202) 4.001 DISP 4.001 DISP Connect the output of the filter to the 8903B HIGH input 4.001 DISP THROUGH ANOTHER 600 ohm load. 4.002 DISP 4.002 DISP Set the UUT FREQUENCY for A 1004HZ output. 4.003 IEEE [@8903]T3T3RR[I] 4.004 MEME 4.005 MEME -65 4.006 MEME 4.007 MEMC -65 -65DB +100U -0U DIST 5.001 DISP CALIBRATION COMPLETE 5.002 END